Warning: The RF safety of the device has not been tested. The patient may only be imaged by landmarking at least 30 cm from the implant, or ensuring the implant is located outside of the RF coil. A patient with a Paltop Dental Implant System/KDG device can be scanned safely in an MR system under the following conditions:
Device Name | Paltop Dental Implant System |
Static Magnetic Field Strength (B0) |
≤ 3.0T |
Maximum Spatial Field Gradient |
20 T/m (2000 gauss/cm) |
RF Excitation |
Circularly Polarized (CP) |
RF Transmit Coil Type |
For body transmit coil, landmarking at least 30 cm from the implant, or ensuring the implant is located outside of the coil. Extremity T/R coils permitted. Excludes Head T/R coil. /p> |
Operating Mode |
Normal Operating Mode in the allowed imaging zone |
Maximum Whole-Body SAR |
2 W/kg (Normal Operating Mode) |
Maximum Head SAR |
Not evaluated for head landmark |
Scan Duration |
No specific constraints due to implant heating |